Utilization of the flow ratio measured by echocardiography (FRecho) compared to the flow ratio by right heart catheterization (FRrhc) for detecting Eisenmenger syndrome in uncorrected acyanotic adult congenital heart disease (ACHD)

Abstract

Background

The increasing number of adult congenital heart disease (ACHD) patients, especially in low- and middle-income countries (LMICs), necessitates effective management methods. The Qp/Qs or flow ratio (FR) is crucial for this purpose because one of the indications for closure is a significant shunt. This study compares the utility of the transthoracic echocardiography (TTE)-derived flow ratio (FRecho) with that of the right heart catheterization (RHC)-derived flow ratio (FRrhc) to guide clinical decisions in resource-limited settings.

Materials and methods

This cross-sectional study in Bandung, Indonesia, included 36 patients with uncorrected acyanotic ACHD who underwent both RHC and TTE on the same day. FRecho was calculated using stroke volumes of the respective ventricles derived from Doppler measurements, and FRrhc was measured using indirect Fick’s method.

Results

Of the 36 patients, 80.6% were female, with a median age of 31 (18–65) years. The majority had secundum atrial septal defects (61.1%). The mean FRecho was 2.8 ± 1.5 and the median FRrhc was 1.69 (0.46–3.89). FRecho showed a significant positive correlation with FRrhc (ρ = 0.656, p < 0.001). Bland‒Altman analysis revealed a mean difference of 1 (-1.4–3.3). Subgroup analysis of patients with a FRrhc shunt < 1 showed a mean difference of 0.7 (-1–2.3).

Conclusion

TTE-derived FRecho tends to overestimate FR compared to FRrhc. FRecho should not be used as a surrogate for FRrhc in this population.

Peer Review reports

Background

The number of patients with adult congenital heart disease (ACHD) is increasing due to advances in diagnostic and therapeutic modalities. Uncorrected congenital heart disease is more common in low- and middle-income countries (LMICs) due to late diagnosis, economic problems, or a lack of education about the importance of defect correction [1].

Acyanotic ACHD represents a spectrum of clinical presentations, ranging from asymptomatic pulmonary arterial hypertension (PAH) to Eisenmenger syndrome (ES), characterized by reversal (pulmonary-to-systemic) or bidirectional shunting due to severe elevation of pulmonary vascular resistance (PVR). The Qp/Qs ratio, or flow ratio (FR), is a tool for quantifying shunt direction and significance. A FR < 1 indicates a dominant right-to-left shunt, while a FR > 1 indicates a dominant left-to-right shunt [2, 3].

Providing optimal care for ACHD patients is challenging, especially in resource-limited settings such as Indonesia [4]. Diagnostic and therapeutic interventions often require a high level of expertise and resources. According to the 2020 ESC Guidelines for the Management of Adult Congenital Heart Disease, the measurement of PVR, and FR measurement is an integral part of the indication for defect closure [5]. Magnetic resonance imaging (MRI) is a good alternative for noninvasively measuring FR. However, radiation exposure, high cost, and machine availability are limitations in developing countries. On the other hand, transthoracic echocardiography (TTE) is a modality that is easy to find and widely used in diagnosing CHD patients and finding other necessary information [6]. Measuring the pressure gradient between ventricles or great arteries could be used to determine the direction of the shunt, especially in VSD or PDA, since the pressure gradient between the defects is high. However, these methods are not applicable in ASD patients, and clinicians often rely on color Doppler to qualitatively determine the direction of the shunt since the pressure gradient between the atria is low [7].

FR can be measured by TTE as the cross-sectional area times the average velocity of the blood passing through the vessel from the right ventricle and the left ventricle [8]. TTE estimation of FR could be used to estimate shunt flow, especially in ASD patients. However, data regarding the measurement of FR by TTE compared to RHC is scarce. In this study, we aimed to explore the utility of TTE in assessing FR compared to RHC for guiding cardiologists in LMICs, especially in low-resource settings, in making clinical decisions.

Materials and methods

This cross-sectional study was conducted in Bandung, Indonesia, in accordance with the Declaration of Helsinki and was approved by the Hasan Sadikin Hospital Ethics Committee, and written informed consent was obtained from each participant. Consecutive patients with uncorrected acyanotic ACHD who underwent RHC procedures and echocardiography on the same day at Hasan Sadikin Hospital Bandung from 2023 to 2024 composed the population. The inclusion criterion in this study was adults ≥ 18 years. We excluded those with complex congenital heart disease, great arteries level shunt, poor echo window, pregnancy, or an LVEF < 50% on echocardiography.

Echocardiography

A Philips EPIQ CVx was used. The flow quantification (Q) by TTE was derived from stroke volume (SV) multiplied by heart rate (HR). Qp is the pulmonic flow, derived from the pulmonic (right ventricle) SV multiplied by the HR, and Qs is the systemic flow, derived from the systemic (left ventricle) SV multiplied by the HR. Furthermore, the equation was simplified as Q = SV of the respective ventricle since HR was consistent between ventricles. FR or Qp/Qs measured by echo (FRecho) was the ratio of RVSV/LVSV.

SV was calculated as the cross-sectional area (CSA) multiplied by the velocity time integral (VTI) of the respective ventricle. RVOT VTI was obtained by placing a pulse wave (PW) Doppler sample within the pulmonic valve from the parasternal short-axis view. The RVOT diameter was obtained at the end diastolic phase, measured at its distal using the inner edge-to-inner edge technique in 2D measurement at the PSAX view. LVOT VTI was obtained by placing the PW Doppler device within the aortic valve from the apical 5-chamber view. LVOT diameter was measured at the same location at which the highest velocity signal was obtained. All the echocardiographic measure were measured in concordance with American Society of Echocardiography (ASE) guidelines [8]. All the acquired data were validated by a certified congenital heart disease cardiologist.

Right heart catheterization

Right heart catheterization was performed using femoral vein access. Continuous measurements of pressure and oxygen saturation were performed. FR was calculated using indirect Fick’s method as Qp/Qs and defined as Eisenmenger when the FR < 1. All examinations were performed by a congenital heart disease specialist as a single observer.

Statistical analysis

The analysis started with descriptive analysis and the normality test of numerical variables in the form of the Kolmogorov‒Smirnov test. Normally distributed data are expressed as the mean ± standard deviation (SD); nonnormally distributed data are expressed as the median (range min – max). Paired t tests or Mann‒Whitney tests were used to determine differences between the FRrhc groups, with a cutoff of 1. Pearson correlation tests or Spearman’s rho tests were used depending on the normality of the data to determine the correlation between FRecho and FRrhc. Bland‒Altman analysis was used to assess bias, with clinically acceptable results of -0.5 and 1.5. If there was good agreement, we analyzed its specificity, sensitivity, negative predictive value, and positive predictive value with a simple 2 × 2 table. The data obtained were recorded in a special form and then processed through the IBM SPSS version 25.0 program for 64-bit Windows.

Results

A total of 36 patients were included in the study; 29 (80.6%) participants were female, with a median age of 31 (18–65) years. Most patients were diagnosed with Secundum ASD (61.1%). The median room air oxygen saturation was 94 (71–99) %. The median RVOT VTI was 16.6 (7–46.2) cm, and the mean RVOT diameter was 30.4 ± 5.7 mm. The mean RV basal height was 50 (38–77) mm. The mean pulmonary artery pressure measured by RHC was 57.5 ± 22.2 mmHg. The median PVR according to the RHC was 10.5 (0.13–33.9) WU. For detailed information, see Table 1.

Table 1 Baseline characteristics

Right heart catheterization was conducted and found a significant difference in all RHC parameters between ES and non-ES. Table 2. Furthermore, patients were grouped based on correctability according to current ESC ACHD guidelines. No significant differences were found in age, BMI, RV basal, TR Vmax, PVAccT, and FRecho between the groups (p = 0.203, p = 0.077, p = 0.095, p = 0.05, p = 0.071, p = 0.105, respectively). The median peripheral oxygen saturation was significantly lower in non-correctable group [92 (71–99) vs. 96 (92–99), p < 0.001]. FRecho was significantly lower in non-correctable group [1.3 (0.46–3.89) vs. 2.97(1.5–3.75), p = 0.002] Table 3.

Table 2 Differences in RHC parameters between groups
Table 3 Differences in parameters between correctability

Spearman’s rho bivariate analysis was performed and revealed a significant positive correlation between FRecho and FRrhc (p < 0.001, ρ = 0.656). Moreover, we make subgroup bivariate analysis based on the pre-tricuspid shunt and post tricuspid shunt. We found significant positive correlation between the FRecho and FRrhc from pre-tricuspid shunt and post tricuspid shunt subgroup (p < 0.001, ρ = 0.594 and p < 0.001, ρ = 0.909, respectively) Fig. 1. Furthermore, we used Bland‒Altman analysis to assess bias. The overall mean difference in the whole population was 1 (-1.4–3.3), indicating good agreement but not meeting the a priori clinically accepted results. Subgroup analysis of patients with a FRrhc shunt < 1 was performed. The results were similar, with a mean difference of 0.7 (-1–2.3). Additionally, we performed subgroup analysis of patients with pre- and post- tricuspid shunts. The results were not clinically accepted, with mean difference of 1.3 (-1–3.5) and 0.3 (-1.7–2.2) respectively Fig. 2.

Fig. 1
figure 1

Relationship between FRecho and FRrhc

Fig. 2
figure 2

Bland‒Altmann analysis between FRecho and FRrhc

Discussion

Accurate measurement of FR is crucial for the management of uncorrected acyanotic adult congenital heart disease. The European Society of Cardiology (ESC) recommends the FR threshold for closure in patients with ASD, VSD, and PDA. Patients who experienced a FR < 1.5 were considered to be at risk of defect closure [6]. Color flow Doppler mapping is usually performed to determine the direction of flow in daily practice. However, this method lacks precision compared to quantitative measurements, for instance, FR calculated using the ratio of right ventricular stroke volume (RVSV) to left ventricular stroke volume (LVSV) [9].

Theoretically, flow (Q) measured by echocardiography is derived from the CSA and the average blood cell velocity through the valve orifice using PW Doppler during the flow period. This measurement is more accurate under a laminar flow with a flat profile. VTI was traced from the outer edge of the brightest portion and averaged over several cardiac cycles, especially in patients with atrial fibrillation [10].

In our study, we found that FRecho was poorly correlated clinically with FRrhc and tended to overestimate the value of FR, despite a good positive correlation between the tests. Therefore, FRecho should not be used to estimate FR in this population. This observation was discordant with some previous findings that the Doppler index of the RVSV/LVSV was clinically useful for estimating the FR, especially in ASD patients [11, 12].

This discrepancy could be attributed to significant regurgitation in the aortic or pulmonic valve [12]. Furthermore, the different positions of the PW Doppler sample can significantly affect the measurements. High cardiac output can also alter RVOT VTI measurements [5].

The last possible explanation for this result is the accuracy of anatomical measurements, especially the RVOT diameter. The American Society of Echocardiography (ASE) reported that the pulmonary annulus is the most difficult area to assess because of poor visualization of the annulus diameter and because the RVOT contracts during systole. In contrast with the RVOT, the LVOT has little variability during systole and therefore has a more precise measurement [10].

Our findings were similar to those of Faherty et al., who reported that FRecho has poor correlation and agreement and tends to overestimate the degree of shunting. In their study, a pediatric population was included, and the time interval between TTE and RHC ranged from 58 days. They hypothesized that the results were due to errors in measuring the anatomy [13]. Similarly, we also hypothesized that the overestimation of FR was due to the complex anatomy of the RVOT and limitations of the acoustic window for adequate quantification of the echo [14]. Estimation of cross-sectional areas in the RVOT was more difficult in the presence of the dilated main pulmonary artery, which is commonly present in pulmonary arterial hypertension [15].

Our study has several limitations. First, we did not perform an interrater reliability test, which might introduce observer bias, especially in RVOT measurements. Our study was also conducted in a single center; therefore, the results may not be generalizable.

Conclusion

Although it is theoretically feasible to measure FR by echocardiography, the results were not clinically acceptable in uncorrected acyanotic ACHD patients. FRecho using the RVSV/LVSV tends to overestimate the FR value and cannot be used as a surrogate parameter of the FRrhc despite its availability.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

ACHD:

Adult Congenital Heart Disease

ASD:

Atrial Septal Defect

AVSD:

Atrioventricular Septal Defect

BMI:

Body Mass Index

ESC:

European Society of Cardiology

FR:

Flow Ratio

FRecho :

Echocardiography-Derived Flow Ratio

FRrhc :

Right Heart Catheterization-Derived Flow Ratio

LMICs:

Low- and Middle-Income Countries

LVOT:

Left Ventricular Outflow Tract

mPAP:

Mean Pulmonary Artery Pressure

PAH:

Pulmonary Arterial Hypertension

PDA:

Patent Ductus Arteriosus

PH:

Pulmonary Hypertension

PVR:

Pulmonary Vascular Resistance

PVAccT:

Pulmonic Valve Acceleration Time

Qp/Qs:

Pulmonary to Systemic Flow Ratio

RHC:

Right Heart Catheterization

RV:

Right Ventricle

RVOT:

Right Ventricular Outflow Tract

SV:

Stroke Volume

SVR:

Systemic Vascular Resistance

TTE:

Transthoracic Echocardiography

TR Vmax:

Tricuspid Regurgitation Maximal Velocity

VSD:

Ventricular Septal Defect

VTI:

Velocity Time Integral

WU:

Wood Units

References

  1. Wamala I, Gongwer R, Doherty-Schmeck K, Jorina M, Betzner A, Zheleva B et al. Infrastructure availability for the care of congenital heart disease patients and its influence on case volume, complexity and access among healthcare institutions in 17 Middle-Income countries. Glob Heart. 2021;16(1):75.

  2. Sommer RJ, Hijazi ZM, Rhodes JF. Pathophysiology of congenital heart disease in the adult: Part I: Shunt lesions. Circulation. 2008;117(8):1090–9.

    Article  PubMed  Google Scholar 

  3. Arvanitaki A, Giannakoulas G, Baumgartner H, Lammers AE. Eisenmenger syndrome: Diagnosis, prognosis and clinical management. Heart. 2020;106(21):1638–45.

    Article  CAS  PubMed  Google Scholar 

  4. Qanitha A, Qalby N, Amir M, Uiterwaal CSPM, Henriques JPS, de Mol BAJM et al. Clinical cardiology in South East Asia: Indonesian lessons from the present toward improvement. Glob Heart. 2022;17(1):66.

  5. Detection of atrial shunt lesions with a single echocardiographic parameter - PMC [Internet]. [cited 2024 Apr 30]. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7297847/

  6. Baumgartner H, De Backer J, Babu-Narayan SV, Budts W, Chessa M, Diller GP, et al. 2020 ESC guidelines for the management of adult congenital heart disease: The task force for the management of adult congenital heart disease of the European society of cardiology (ESC). This study was endorsed by the association for European pediatric and congenital cardiology (AEPC) and the international society for adult congenital heart disease (ISACHD). Eur Heart J. 2021;42(6):563–645.

    Article  CAS  PubMed  Google Scholar 

  7. Awasthy N, Radhakrishnan S. Stepwise evaluation of left to right shunts by echocardiography. Indian Heart J. 2013;65(2):201–18.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Zoghbi WA, Adams D, Bonow RO, Enriquez-Sarano M, Foster E, Grayburn PA, et al. Recommendations for noninvasive evaluation of native valvular regurgitation: A report from the American society of echocardiography developed in collaboration with the society for cardiovascular magnetic resonance. J Am Soc Echocardiogr. 2017;30(4):303–71.

    Article  PubMed  Google Scholar 

  9. Silvestry FE, Cohen MS, Armsby LB, Burkule NJ, Fleishman CE, Hijazi ZM, et al. Guidelines for the echocardiographic assessment of atrial septal defect and patent foramen ovale: From the American society of echocardiography and society for cardiac angiography and interventions. J Am Soc Echocardiogr. 2015;28(8):910–58.

    Article  PubMed  Google Scholar 

  10. Recommendations for Quantification of Doppler Echocardiography [Internet]. [cited 2024 May 6]. Available from: https://www.asecho.org/guideline/recommendations-for-quantification-of-doppler-echocardiography/

  11. Kitabatake A, Inoue M, Asao M, Ito H, Masuyama T, Tanouchi J, et al. Noninvasive evaluation of the ratio of pulmonary to systemic flow in atrial septal defect by duplex doppler echocardiography. Circulation. 1984;69(1):73–9.

    Article  CAS  PubMed  Google Scholar 

  12. Sanders SP, Yeager S, Williams RG. Measurement of systemic and pulmonary blood flow and QP/QS ratio using doppler and two-dimensional echocardiography. Am J Cardiol. 1983;51(6):952–6.

    Article  CAS  PubMed  Google Scholar 

  13. Faherty E, Rajagopal H, Lee S, Love B, Srivastava S, Parness IA, et al. Correlation of transthoracic echocardiography-derived pulmonary to systemic flow ratio with hemodynamically estimated left to right shunt in atrial septal defects. Ann Pediatr Cardiol. 2022;15(1):20–6.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Marín Rodríguez C, Sánchez Alegre ML, Lancharro Zapata Á, Alarcón Rodríguez J. What radiologists need to know about the pulmonary–systemic flow ratio (Qp/Qs): What it is, how to calculate it, and what it is for. Radiologia. 2015;57(5):369–79.

    Article  PubMed  Google Scholar 

  15. Stern H, Vogel M, Emmrich P, Bühlmeyer K. The accuracy of doppler echocardiography determination of shunt volume in children with isolated atrial septal defect and left-right shunt. Z Kardiol. 1989;78(10):674–9.

    CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Open access funding provided by University of Padjadjaran

This research was not funded.

Author information

Authors and Affiliations

Authors

Contributions

CJC and AFK made the concept and the framework. CJC, AFK, and NS wrote the manuscript. CJC, and NS made the analysis. All authors approved and reviewed the manuscript.

Corresponding author

Correspondence to Charlotte Johanna Cool.

Ethics declarations

Ethics approval and consent to participate

The research was performed in accordance with the Declaration of Helsinki and was approved by the Hasan Sadikin Hospital Ethics Committee.

Informed consent

Written informed consent was obtained from each participant.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cool, C.J., Khalid, A.F. & Sukmadi, N. Utilization of the flow ratio measured by echocardiography (FRecho) compared to the flow ratio by right heart catheterization (FRrhc) for detecting Eisenmenger syndrome in uncorrected acyanotic adult congenital heart disease (ACHD). BMC Med Imaging 25, 136 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12880-025-01651-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12880-025-01651-y

Keywords